Review Articles

A review of hemodynamic response to the use of different types of laryngoscopes

Roopa Sachidananda, DA, DNB1, Umesh G, MD2, Safiya I. Shaikh, MD3

1Assistant professor; 3Professor and HoD

Department of Anaesthesiology, Karnataka Institute of Medical Sciences, Hubli, Karnataka 580 021, (India)

2Professor and Head, Department of  Anaesthesiology, Dharwad Institute of Mental Health &Neuro Sciences, Dharwad  & Karnataka Institute of Medical Sciences, Karnataka, (India)

Correspondence: Dr Roopa Sachidananda, Karnataka Institute of Medical Sciences, P B Road, Vidyanagar, Hubli, Karnataka 580022, (India); Phone: +919448658220; E-mail: roopasac@rediffmail.com 

ABSTRACT

Endotracheal intubation is the gold standard in airway management either as a means to administer general anesthesia or for mechanical ventilation in critically ill patients for protecting the airway when the airway reflexes are dysfunctional. Macintosh laryngoscope is the standard laryngoscope used for intubation though advances in science have resulted in development of laryngoscopes of different designs. The process of laryngoscopy and intubation (L&I) can result in significant hemodynamic response and this is a topic of debate and research in anesthesia. A comprehensive review of hemodynamic responses to various laryngoscope designs has been undertaken here. Tracheal intubation contributes to more hemodynamic response compared to laryngoscopy alone. The hemodynamic responses to L&I are exaggerated in the elderly and those with uncontrolled hypertension. Orotracheal intubation causes less hemodynamic response as compared to nasotracheal intubation. Laryngoscope design, duration of L&I and the forces applied on the laryngoscope all contribute to hemodynamic fluctuations. McCoy blade and videolaryngoscopes where L&I can be performed without the aid of stylets provide better attenuation of hemodynamic response compared to intubation using the Macintosh laryngoscope. Fibreoptic orotracheal intubation with the aid of combined lingual traction and jaw thrust maneuver provides superior attenuation of hemodynamic response compared to use of laryngoscope.

Key words: Airway Management; Laryngoscopy; Laryngoscopes; Intubation; Intratracheal; Cardiovascular Physiological Phenomena; Hemodynamics; Stress; Physiological; Physiological Stress Reaction

Citation: Sachidananda R, Umesh G, Shaikh SI. A review of hemodynamic response to the use of different types of laryngoscopes. Anaesth Pain & Intensive Care 2016;20(2):–

Received: 14 February 2016; Reviewed: 26 March 2016; Corrected: 10 March, 26 April, 9 June 2016; Accepted: 9 June 2016

INTRODUCTION

Endotracheal intubation (ETI) is well recognized to be the gold standard in airway management to administer general anesthesia (GA). The indications for mechanical ventilation in critically ill patients include relieving airway obstruction, respiratory failure, prevent aspiration, airway protection, etc. The Macintosh laryngoscope is the standard laryngoscope used for ETI though advances in science have resulted in the development of several laryngoscopes of different designs.1,2,3 The process of laryngoscopy and intubation (L&I) can result in significant hemodynamic response and therefore, limiting or taming this response is a topic of debate and research in anesthesia.

The hemodynamic response to L&I is regulated by the hypothalamo-pituitary-adrenocortical and sympathetic adreno-medullary response.4 As a result of which there is secretion of cortisol, norepinephrine and epinephrine. The consequence of this neuro–endocrine system may vary from milder problems such as tachycardia, hypertension and occasional dysrhythmias to life threatening problems such as angina, myocardial infarction, stroke, etc. The hemodynamic response to L&I was first enunciated by King et al in 1951, although ETI was being practiced since its inception into anesthetic practice by Rowbotham and Magil in 1921.5,6

Prior to the invention of muscle relaxants, the hemodynamic response to L&I was little known, probably because the patients were intubated under deep cyclopropane or ether anesthesia and may be their cardiovascular responses were obtunded due to much deeper planes of anesthesia that were necessary to render the patient immobile during the surgical procedure in the absence of neuromuscular blockade. Further, the monitoring for the hemodynamic response also was suboptimal. Shortly after the introduction of d-tubocurarine, there was a considerable change in the scenario as the hemodynamic response to L&I started becoming obvious due to considerable decrease in the depth of the anesthesia with regular use of neuromuscular blockade. It was also recognized that simply increasing the anesthetic depth afforded considerable protection from these responses.6

Several drugs have been used, different laryngoscope designs have evolved and anesthetic techniques have been modified to attenuate these reflexes. A comprehensive review of hemodynamic responses to various laryngoscope designs has been undertaken here.

Literature search: Systemic search of literature was done using key words; laryngoscopy response; ETI and response; hemodynamic response and intubation; sympathetic response and laryngoscopy or intubation; laryngoscopes and hemodynamics; and pressor response and laryngoscopy or intubation. We have also attempted to search articles using names of various types of laryngoscopes in association with hemodynamic response/pressor response/intubation response.

Hemodynamic response and the receptor location:

Mechanical stimulation of the upper respiratory tract, mainly: the nose, the epipharynx, and the tracheobronchial tree induce reflex cardiovascular responses associated with enhanced neuronal activity in cervical sympathetic efferent fibres. While stimulation of the epipharynx elicits maximum response, tracheobronchial tree elicits least response.7 Cardiovascular response to ETI is initiated by glossopharyngeal nerve (stimulus superior to anterior surface of epiglottis) and by vagus nerve (stimulus below posterior epiglottis down into the lower airway). Hemodynamic response to L&I results in diffuse autonomic response with a widespread release of norepinephrine from adrenergic nerve terminals and secretion of epinephrine from adrenal medulla along with activation of the renin angiotensin system.8

Physiology of hemodynamic response:

This is maximum at approximately 30-45 seconds after L&I.9 Blood Pressure (BP), Heart Rate (HR), plasma adrenaline, noradrenaline and vasopressin concentrations increase slightly in response to L&I; all returning to baseline within 5 min with no change in angiotensin converting enzyme activity in normotensives. However, a three-fold increase in plasma noradrenaline levels which returned to baseline nearly 10 minutes following L&I was observed in hypertensives. Further, an increase in plasma adrenaline level was observed in hypertensives 1 minute after laryngoscopy. 10,11,12,13

Effects of hemodynamic responses on organ systems:

A rate pressure product (RPP) in excess of 11000 has been associated with signs of myocardial ischemia in patients with coronary artery disease.14 Sudden increase in blood pressure may cause rupture of aortic / cerebral aneurysm, increase cerebral blood flow due to increased cerebral metabolic activity and systemic cardiovascular effects, dysrhythmias, transient increase in choroid blood flow which can force vitreous gel forward into the anterior chamber during open eye surgery or can increase intraocular pressure in an intact eye15. The normal autoregulation mechanism may not be effective because of disease. Patients with raised intracranial pressure who have minimal reserve in intracranial compliance are at a risk for brainstem herniation and sudden death during L&I16.

Hemodynamic response in relation to age

In infants and small children, response may manifest initially as bradycardia owing to an increased vagal tone.8 In geriatric patients, SBP and MAP increased significantly though the tachycardia response was less severe as the age advanced which was attributed to impaired β response with normal α responsiveness. It was also noted that the mean plasma norepinephrine concentrations were significantly less in the elderly. 17

Effect of force applied, duration and number of attempts

Various forces applied onto a Macintosh laryngoscope (the forces acting along the axis of the handle, as well as forces exerted by anesthetist to prevent the laryngoscope from turning) were studied for their influence on hemodynamics. A modified laryngoscope was designed to measure these forces and four different parameters were determined:

  1. The duration of laryngoscopy
  2. The maximally applied forces
  3. The mean force
  4. The integral of the force over the time (area under the curve)

The duration of laryngoscopy, the forces applied parallel to the axis of the handle and the increased stretching of the tissues were found to be responsible for the hemodynamic changes.18,19

HEMODYNAMIC RESPONSE TO VARIOUS LARYNGOSCOPES (TABLE 1):

Miller and McCoy laryngoscope

When the stress response during laryngoscopy without intubation was compared among Macintosh, Miller and McCoy laryngoscopes, the maximum response was obtained with the use of Miller and minimum response with the McCoy laryngoscope. The tip of the Miller’s blade which is inserted posterior to the epiglottis stimulating the vagus, causes maximum response.20

In order to get a superior glottic view at first attempt itself, the McCoy laryngoscope was designed. The design of the blade of this laryngoscope is such that the tip of the blade though rigid is mobile which can be lifted up using a lever available at the handle of the laryngoscope. When a poor view of glottis is obtained despite best efforts at laryngoscopy, the laryngoscopist presses this lever which results in the epiglottis being lifted away due to upward movement of the tip of the laryngoscope, contributing to better view of the glottis. The stress response to laryngoscopy appears to be less marked with the McCoy blade probably due to a reduction in the force necessary to obtain a clear view of the larynx.21,22

 Table 1: Hemodynamic response and various laryngoscopes

Serial No. Author

Year

Country

Design Sample size Airway Results Reasons cited
1 McCoy EP et al

199522

Belfast

RCT Macintosh N=10

McCoy N=10

Stress response to laryngoscope with McCoy was less compared to Macintosh McCoy requires less force than Macintosh for visualization of larynx
2 Garg S et al

200323

India

RCT Macintosh

N=40

Intubating LMA

N=40

 

Patients with restricted mouth opening were excluded Less pressor response on ETT insertion using ILMA compared to the conventional laryngoscope
3 Sener EB et al

201224

Turkey

RCT ILMA

N=21

Macintosh

N=21

Patients with difficult intubation were excluded The SBP, DBP and RPP was higher in ILMA group The intense and repeated oropharyngeal and tracheal stimulation resulting from ILMA induces greater pressor response
4 Kavitha J et al

201125

India

RCT ILMA

N=30

Macintosh

N=30

ILMA offers no advantage in attenuating the hemodynamic response compared to direct laryngoscope Direct laryngoscopy is comparatively a faster method to secure tracheal intubation when compared to ILMA
5 Aghdaii et al

201228

Iran

RCT Direct laryngoscope

N=25

Fibreoptic scope

N=25

Fibreoptic intubation has no advantage in attenuating hemodynamic response to intubation in patients undergoing CABG
6 Pournajafian AR et al

201434

Tehran

RCT Macintosh

N=53

GlideScope®VLS

N=53

Airway up to Mallampati class III No significant difference between the two groups in hemodynamic response Longer intubation time using GlideScope®VLS
7 Dashti et al

201435

Iran

RCT GlideScope®VLS

N=30

Macintosh

N=29

 

Normal airway Hemodynamic changes during tracheal intubation were less significant with GlideScope® than Macintosh Upward lifting force required to expose the glottis is much less with GlideScope®
8 Woo et al

201243

Korea

 

RCT Pentax AWS®VLS

N=50

Macintosh

N=50

Burns patients up to Mallampati class III Less increase in hemodynamic response with Pentax-AWS®VLS compared to Macintosh Pentax AWS®VLS requires less force to lift structures in oral cavity to expose the glottis
9 Heeseung Lee et al

201342

Korea

RCT Pentax AWS®

N=20

Macintosh

N=20

Normal anatomy Pentax AWS® provides greater hemodynamic stability during tracheal intubation compared to Macintosh Pentax AWS® causes minimal displacement of the tongue and other soft tissue of airway structures
10 Ko et al

201245

Korea

RCT OptiscopeTM

N=30

Macintosh

N=30

Mallampati class IV excluded OptiscopeTM does not modify hemodynamic response. OptiscopeTM provides better view of the vocal cords though there is no benefit on hemodynamic response
11 Koh et al

201047

Seoul

RCT Airtraq

N=25

Macintosh N=25

Mallampati class IV were excluded  Hemodynamic changes were not significantly different between the two groups
12 Kitamura et al

200149

Tokyo

RCT StyletScope N=13

Macintosh N=11

StyletScope caused less hemodynamic response compared to Macintosh The stimulus from direct laryngoscopy is eliminated
13 Kanchi et al

201144

India

RCT Macintosh

N=15

Pentax AWS VLS

No difference in the hemodynamic response Time taken for intubation using VLS was longer
14 Maassen RL et al

201257

Netherlands

RCT – cross over VLS Vs Conventional laryngoscope

 

 

Hemodynamic responses during endotracheal intubation using VLS was less compared to direct laryngoscopy

RCT=Randomized control trial; VLS= Videolaryngoscope;N=number; ETT= endotracheal tube; LMA= laryngeal mask airway; ILMA= intubating laryngeal mask airway; SBP=systolic blood pressure; DBP=diastolic blood pressure; RPP= rate pressure product; CABG=coronary artery bypass graft; AWS= airway scope

Intubating laryngeal mask airway (ILMA)

ILMA guided orotracheal intubation does not distort the base of the tongue or directly stimulate the receptors of the larynx. In patients undergoing coronary artery surgery, a significant decrease in mean norepinephrine concentration was observed with both ILMA guided intubation and L&I. However, this drop was more marked in the patients intubated via ILMA than in patients undergoing direct laryngoscopy. RPP showed similar response. Similar results were obtained by Garg et al. But the trial conducted by Sener et al failed to attenuate the hemodynamic response with the use of ILMA, where, in contrary, there was higher response. The longer duration, repeated airway manipulation, stimulation of supralaryngeal area which is rich in nociceptive receptors, removal of ILMA and using stabilizing rod to advance the tracheal tube may have induced greater pressor response in these patients. 23,24,25

Fibreoptic intubation:

The fibreoptic endoscope enables the visualization of the glottis with minimum force, but the need of maneuvers or instruments to clear the airway result in hemodynamic response.

Hemodynamic response was found to be more severe with the use of Macintosh laryngoscopy as an aid for fibreoptic orotracheal intubation as compared to a combined maneuver of lingual traction and jaw thrust. Factors like prolonged tracheal stimulation, jaw thrust maneuver in fibreoptic intubation and external neck manipulation also contribute to the hemodynamic response. Effective suppression of hemodynamic response was achieved by combining fibreoptic intubation with 4% lidocaine spray of the larynx and glossopharyngeal nerve block. 26,27 But the study by Aghdaii et al. showed no advantage with the use of fibreoptic scope for intubation in patients undergoing coronary artery bypass graft (CABG).28

Nasotracheal intubation:

The hemodynamic response after nasotracheal intubation is significantly greater than that after orotracheal intubation as there would be considerable stimulation of the irritant receptors in the nostrils.29

Videolaryngoscopes (VLS)

The VLS have been introduced in 2002 to aid the intubation without the need for the alignment of oral, pharyngeal and laryngeal axes. In most of the VLS the blades have been angulated to enhance the visualization of the glottis. This necessitates the use of stylet or laryngeal manipulation to aid the successful intubation. The use of stylet or laryngeal manipulations has been shown to increase the hemodynamic response.

Glidescope

The Glidescope that consists of a blade with 60º curvature and a handle in one piece does not require alignment of oropharyngeal axis to visualize the vocal chords. The laryngoscope is devised in such a way that less upward lifting force is required during L&I thereby reducing the stimulus on to the oropharyngeal structures.30 However, some studies failed to show the advantage of Glidescope compared to Macintosh in attenuating stress response to intubation as manipulation due to stylet in Glideview laryngoscopy resulted in higher stimulus offsetting the advantage.31,32 When the technique of intubation was standardized with the use of stylet in both groups, the hemodynamic response was found to be similar in Glidescope, trachlight and direct laryngoscopy groups. Authors opined that the hemodynamic response was mainly because of tracheal irritation rather than laryngeal irritation.33 In studies where the hemodynamic response was more in the Glidescope compared to Macintosh, the cause was attributed to either longer time of intubation or as a result of less expertize or due to response to endotracheal tube in trachea.34 Glidescope VLS is a better option in patients with untreated hypertension.35 The hemodynamic response to intubation were similar with both the Glidescope and the McGrath VLS.36

Pentax-AWS

It is a VLS with anatomically shaped blade provides high resolution image of the glottis.37 It requires considerably less lifting force during laryngoscopy with this device and hence it causes less movement of the upper cervical spine with minimum displacement of the tongue and other soft tissues of the airway structures.38,39,40,41 Pentax– AWS provides greater hemodynamic stability than Macintosh blade laryngoscope and seems to be a safe alternative especially for patients susceptible to hemodynamic stress and for patients with severe burns.42,43 The use of Pentax AWS did not provide any benefit in attenuating hemodynamic response to L&I in patients undergoing primary coronary artery bypass graft surgery.44

Optiscope

It is a semi rigid fiberscope with adjustable angle tip. Although ETI performed using Optiscope gave a better view of vocal cords, it did not modify the hemodynamic response.45

Airtraq ® optical laryngoscope

It is a single use rigid VLS that has been used to intubate patients with normal and difficult airways. It incorporates video or fibreoptic imaging into the blade or stylet and hence provides an improved glottis view without the need for alignment of oral, pharyngeal and laryngeal axes. The Airtraq ®optical laryngoscope resulted in clinically negligible rise in heart rate while the blood pressure was not affected. 46Airtraq® also produces a superior laryngeal view over the Macintosh laryngoscope in patients with Philadelphia collar without any difference in hemodynamics or intubation time.47

StyletScope

StyletScope is a fibreoptic intubation device with a lightweight stylet, maneuverability of the distal tip, a built in light source. Distal tip of the stylet, together with the ETT, can be flexed 75 degree anteriorly. The intubation with StyletScope does not require direct laryngoscopy. Hence the Stylet Scope caused less hemodynamic response compared to Macintosh laryngoscope.48, 49

C-MAC VLS

This VLS has an original standard Macintosh blade shape, but with a flat proximal finish. Force applied on the maxillary teeth is minimum. The less force applied to lift the epiglottis should theoretically cause less hemodynamic response. However the study by Sarkilar et al. showed similar response in both C-MAC and Macintosh intubation. The author attributed the results to the use of various anesthetic agents. 50

Truview VLS

Truview EVO2 laryngoscope has an integrated optical lens system with unique angulated blade which provides optimal indirect view of the glottis. The hemodynamic response was less with Truview PCD when compared with Macintosh in the study by Bag et al.51

McGrath Series 5 VLS

McGrath Series 5 has a video camera with angulated blade and a light source at the end of the blade. Although this VLS provides better glottis view compared to the Macintosh laryngoscope, there was significant hemodynamic response to L & I with both Macintosh and McGrath Series in the study by Arici et al. in obstetric patients. 52 The study by Gupta et al. comparing Macintosh with McGrath VLS did not show any benefit with the use of McGrath VLS in attenuating the hemodynamic response in patients with normal airway.53

When the hemodynamic response in patients with normal airway, undergoing coronary artery bypass grafting (CABG) were compared among Truview PCDTM, McGrath® and Macintosh, the VLS did not provide any advantage over the Macintosh laryngoscope. The VLS is usually introduced into the center of the mouth unlike the conventional Macintosh laryngoscope. This reduces the space available to insert the ETT, necessitating additional maneuvers to negotiate the tube. This also prolongs the  time for intubation. These factors might neutralize the advantages provided by the VLS such as less lifting force needed to obtain a good view of the glottis.54

However the hemodynamic response with GlideScope VLS showed promising results when compared with fibreoptic scope and Macintosh laryngoscope.55

In the study by Tsai et al. where they compared the hemodynamic response of the Airway Scope®, Glidescope and Macintosh laryngoscopes, the response was less with the Airway Scope®. Although the intubation time was significantly longer with Airway Scope® and Glidescope, when compared to the Macintosh, the authors attributed the result to the more invasiveness of the Glidescope laryngoscope. 56

Optical stylets

Various optical stylets have been used to ease intubation. Rigid optical stylets include bonfils, Levitans FPS where the operator can look through the ocular, like in a fiberscope and observe the entrance to larynx. Bonfils intubation endoscope is a semi-rigid optical stylet, 40 cm in length with a fixed anterior curvature of its tip of 40º. Flexible fibreoptic intubation and the bonfils rigid endoscope guided intubation require a similar time for successful orotracheal intubation and cause a similar magnitude of hemodynamic response in female adults under general anesthesia.58

The hemodynamic response to intubation using the Levitan FPS device was greater than that of the Lary-Flex rigid videolaryngoscope. The authors implicated this to the use of Macintosh laryngoscope while using the Levitan FPS device. 59

CONCLUSION

Tracheal intubation contributes to more hemodynamic response compared to laryngoscopy alone. The hemodynamic responses to L&I are exaggerated in the elderly and those with uncontrolled hypertension. Orotracheal intubation causes less hemodynamic response as compared to nasotracheal intubation. Laryngoscope design, duration of L&I and the forces applied on the laryngoscope all contribute to hemodynamic fluctuations. McCoy blade and videolaryngoscopes where laryngoscopy and intubation can be performed without the aid of stylets provide better attenuation of hemodynamic response compared to intubation using the Macintosh laryngoscope. Fibreoptic orotracheal intubation with the aid of combined lingual traction and jaw thrust maneuver provides superior attenuation of hemodynamic response compared to the use of laryngoscope.

Conflict of interest: None declared by the authors

Authors’ Contribution: RS-data collection, data analysis, manuscript preparation, editing; UG- data analysis, manuscript preparation, editing; SIS-editing

REFERENCES

  1. Levitan RM, Heitz JW, Sweeney M, Cooper RM. The complexities of tracheal intubation with direct laryngoscopy and alternative intubation devices. Ann Emerg Med 2011;57:240-247. [PubMed] doi: 10.1016/j.annemergmed.2010.05.035
  2. Savoldelli GL, Schiffer E, Abegg C, Baeriswyl V, Clergue F, Waeber JL. Comparison of the Glidescope, the McGrath, the Airtraq and the Macintosh laryngoscopes in simulated difficult airways. Anaesthesia 2008;63:1358-1364. [PubMed] [Free full text] doi: 10.1111/j.1365-2044.2008.05653.x
  3. Stout DM, Bishop MJ, Dwersteg JF, Cullen BF. Correlation of endotracheal tube size with sore throat and hoarseness following general anesthesia. Anesthesiology 1987;67:419-421. [PubMed] [Free full text]
  4. Schommer NC, Helhammer DH, Kirschbaum C. Dissociation between reactivity of the hypothalamus-pituitary-adrenal axis and the sympathetic-adrenal-medullary system to repeated psychosocial stress. Psychosom Med 2003;65(3):450-460 [PubMed] [Free full text]
  5. King BD, Harris LC, Greifenstein FE, Elder JD, Dripps RD. Reflex circulatory responses to direct laryngoscopy and tracheal intubation performed during general anesthesia. Anesthesiology 1951;12:556-566. [PubMed] [Free full text]
  6. Derbyshire DR, Smith G. Sympathoadrenal responses to anaesthesia and surgery. Br J of Anaesth 1984;56:725-739. [PubMed] [Free full text]
  7. Tomori Z, Widdicombe JG. Muscular, bronchomotor and cardiovascular reflexes elicited by mechanical stimulation of the respiratory tract. J Physiol 1969;200:25a-49a. [PubMed] [Free full text]
  8. Bishop M J, Robert F Bedford, Hae K Kil. Physiologic and pathophysiologic responses to intubation (1996). Jonathan L Benumof. Airway management, Principles and Practice. Missouri, USA: Mosloy P. 102 – 117.
  9. Wycoff C. Endotracheal intubation;effects on blood pressure and pulse rate. Anaesthesiology 1960;21:153-157. [PubMed] [Free full text]
  10. Corbett JL, Kerr JH, Prys-Roberts C, Cramptonsmith A. Cardiovascular disturbances in severe tetanus due to over activity of the sympathetic nervous system. Anaesthesia 1969;24:198-210. [PubMed] [Free full text]
  11. Russel WJ, Morris RG, Frewin DB, Drew SE. Changes in plasma catecholamine concentrations during endotracheal intubation. Br J Anaesth 1981;53:37-39 [PubMed] [Free full text]
  12. Kayhan Z, Aldemir D, Mutlu H, Ogus E. Which is responsible for the hemodynamic response due to laryngoscopy and endotracheal intubation? Catecholamines, vasopressin or angiotensin? Eur J Anaesthesiol 2005;22:780-785.[PubMed] [Free full text]
  13. Low J M, Harey J T, Prys Roberts C, Dagino J. Studies of anaesthesia in relation to hypertension VII: Adrenergic responses to laryngoscopy. Br J Anaesth 1986;58;471-477. [PubMed] [Free full text] doi: 10.1093/bja/58.5.471
  14. Roy WL, Edelist G, Gilbert B. Myocardial ischemia during non-cardiac surgical procedures in patients with coronary – artery disease. Anesthesiology 1979;51:393-397. [PubMed] [Free full text]
  15. Gal TJ. Airway management In:Miller RD, ed. Miller’s Anesthesia 6th Philadelphia: Elsevier Churchill Livingstone, 2005. p1617-1652.
  16. Donlon JV, Doyle DJ, Feldman MA. Anesthesia for eye, ear, nose and throat surgery In:Miller RD, ed. Miller’s Anesthesia 6th Philadelphia: Elsevier Churchill livingstone, 2005. 2527-2549.
  17. Bullington J, Mouton Perry SM, Rigby J, Pinkerton M, Rogers D, Lewis TC, et al.The effect of advancing age on the sympathetic response to laryngoscopy and tracheal intubation. Anesthe Analg 1989;68:603-608. [PubMed]
  18. Bucx MJL, Scheck PAE, Van Geel RTM, Den Ouden AH, Niesing R. Measurement of forces during laryngoscopy. Anaesthesia 1992;47:348-351. [PubMed] [Free full text]
  19. Bucx MJ, Van Geel RT, Scheck PA, Stijnen T. Cardiovascular effects of forces applied during laryngoscopy. Anaesthesia 1992;47:1029-1033. [PubMed] [Free full text] DOI: 10.1111/j.1365-2044.1992.tb04195.x
  20. Nishiyama T, Higashizawa T, Bito H, Konishi A, Sakai T. Which laryngoscope is the most stressful in laryngoscopy; Macintosh, Miller, or McCoy? Masui 1997;46(11) 1519-1524. [PubMed]
  21. Haidry MA, Khan FA. Comparison of hemodynamic response to tracheal intubation with Macintosh and McCoy laryngoscopes. J Anaesthesiol Clin Pharmacol. 2013;29:196-199. [PubMed] [Free full text] doi: 10.4103/0970-9185.111710.
  22. McCoy EP, Mirakhur RK, and McCloskey BV. A comparison of the stress response to laryngoscopy: the Macintosh versus the McCoy blade. Anaesthesia 1995;50:943-946. [PubMed] [Free full text] DOI: 10.1111/j.1365-2044.1995.tb05924.x
  23. Garg S, Sodhi BK, Mohindra BK, Tewari A, Bhambri V, Sharma S. Changes in the haemodynamic responses : Conventional endotracheal intubation Vs intubation with intubating LMA. J Anaesth Clin Pharmacol 2003;19 (3):289-294.
  24. Sener EB, Ustun E, Ustun B, Sarihasan B. Hemodynamic responses and upper airway morbidity following tracheal intubation in patients with hypertension: Conventional laryngoscopy versus intubating laryngeal mask airway. Clinics 2012;67(1):49-54. [PubMed] [Free full text] doi:  10.6061/clinics/2012(01)08
  25. Kavitha J, Tripathy DK, Mishra SK, Mishra G, Chandrasekhar LJ, Ezhilarasu. Intubating condition, hemodynamic parameters and upper airway morbidity: A comparison of intubating laryngeal mask airway with standard direct laryngoscopy. Anesth Essays Res 2011;5:48-56. [Free full text] doi:  10.4103/0259-1162.84190
  26. Venkatesh HK, Umamaheshwara Rao GS. Optimal combination of laryngoscopy and airway anaesthesia to minimize cardiovascular response to tracheal intubation. Journal Anaesth Clin Pharmacol 2004;20:379-386.
  27. Tong JL. Ashworth DR, Smith JE. Cardiovascular responses following laryngoscope assisted, fibreoptic orotracheal intubation. Anaesthesia 2005;60:754-758. [Free full text] DOI: 10.1111/j.1365-2044.2005.04238.x
  28. Aghdaii N, Azarfarin R, Yazdanian F, Faritus SZ. Cardiovascular responses to orotracheal intubation in patients undergoing coronary artery bypass grafting surgery. Comparing fiberoptic bronchoscopy with direct laryngoscopy. Middle East J Anesthesiol 2010;20(6):833-838. [PubMed]
  29. Singh S, Smith JE. Cardiovascular changes after the three stages of nasotracheal intubation. Br J Anaesth 2003;91:667-671. [PubMed] [Free full text] doi: 10.1093/bja/aeg240
  30. Sun DA, Warriner CB, Parsons DG, Klein R, Umedaly HS, Mount M. The Glidescope Video Laryngoscope: randomized clinical trial in 200 patients. Br J Anaesth 2005;94:381-4. [PubMed] [Free full text] doi: 10.1093/bja/aei041
  31. Xue FS, Zhang GH, Li XY, Sun HT, Li P, Li CW, et al. Comparison of hemodynamic responses to orotracheal intubation with the GlideScope videolaryngoscope and the Macintosh direct laryngoscope. J Clin Anesth 2007;19:245-250. [PubMed]
  32. Malik MA, Maharaj CH, Harte BH, Laffey JG. Comparison of Macintosh, Trueview EVO2, Glidescope and Airwayscope laryngoscope use in patients with cervical spine immobilization. Br J Anaesth 2008;101:723-730. [PubMed] [Free full text] doi: 10.1093/bja/aen231.
  33. Siddiqui N, Katznelson R, Friedman Z, Heart rate/blood pressure response and airway morbidity following tracheal intubation with direct laryngoscopy,GlideScope and Trachlight: a randomized control trial. Eur J Anaesthesiol 2009;26(9):740-745. [PubMed] doi: 10.1097/EJA.0b013e32832b138d.
  34. Pournajafian AR, Ghodraty MR, Faiz SHR, Rahimzadeh P, Goodarzynejad H, Dogmehchi E. Comparing Glidescope video laryngoscope and Macintosh laryngoscope regarding hemodynamic responses during orotracheal intubation: a randomized controlled trial. Irans Red Crescent Med J 2014;16 (4):e12334 [PubMed] [Free full text] doi: 10.5812/ircmj.12334
  35. Dashti M, Amini S, Azarfarin R, Totonchi Z, Hatami M. Hemodynamic changes following endotracheal intubation with Glidescope® video- laryngoscope in patients with untreated hypertension. Res Cardiovasc Med 2014 May;3(2):e 17598. [PubMed] [Free full text] doi: 10.5812/cardiovascmed.17598.
  36. Jeon WJ, Kim KH, Yeom JH, Bang MR, Hong JB,Cho SY. A comparison of the Glidescope® to the McGrath® videolaryngoscope in patients. Korean J Anesthesiol 2011;61(1):19-23. [PubMed] [Free full text]  doi: 10.4097/kjae.2011.61.1.19
  37. Suzuki A, Toyama Y, Katsumi, Kunisawa R, Sasaki S Hirota k et al. The Pentax-AWS(R) rigid indirect videolaryngoscope: clinical assessment of performance in 320 cases. Anaesthesia 2008;63:641-647. [PubMed] [Free full text] doi: 10.1111/j.1365-2044.2008.05452.x.
  38. Enomoto Y, Asai T, Kamishima K, Okuda Y. Pentax-AWS, a new videolaryngoscope, is more effective than the Macintosh laryngoscope for tracheal intubation in patients with restricted neck movements: a randomized comparative study. Br J Anaesth 2008;100:544-548. [PubMed] [Free full text] doi: 10.1093/bja/aen002
  39. Malik MA, Maharaj CH, Harte BH, Laffey JC. Comparison of Macintosh, Truview EVO2, Glidescope and Airwayscope laryngoscope use in patients with cervical spine immobilization. Br J Anaesth 2008;101:723-730. [PubMed] [Free full text] doi: 10.1093/bja/aen231
  40. Malik MA, Subramaniam R, Churasia S, Maharaj CH, Harte BH, Laffey JG. Tracheal intubationin patients with cervical spine immobilization: a comparison of the Airwayscope, LMA C trach and the Macintosh laryngoscopes. Br J Anaesth 2009;102:654-661. [PubMed] [Free full text] doi: 10.1093/bja/aep056
  41. Koyama J, Aoyaa T, Kusano Y, Seguchi T, Kawagishi K, Iwashita T et al. Description and first clinical application of Airway Scope for tracheal intubation. J Neurosurg Anesthesiol 2006;18:247-250. [PubMed]
  42. Lee H. The Pentax airway scope versus the Macintosh laryngoscope: Comparison of hemodynamic responses and concentrations of plasma norepinephrine to tracheal intubation. Korean J Anesthesiol 2013;64(4):315-320. [PubMed] [Free full text] doi: 10.4097/kjae.2013.64.4.315.
  43. Woo CH, Kim SH, Park JY, Bae JY, Kwak IS, Mun SH, et al. Macintosh laryngoscope vs pentax-AWS video laryngoscope: comparison of efficacy and cardiovascular responses to tracheal intubation in major burns patients. Korean J Anesthesiol 2012;62(2):119-124. [PubMed] [Free full text] doi: 10.4097/kjae.2012.62.2.119.
  44. Kanchi M, Nair HC, Banakal S, Murthy K, Murugesan C. Haemodynamic response to endotracheal intubation in coronary artery disease: Direct versus videolaryngoscopy. Indian J Anaesth 2011;55:260-265. [PubMed] [Free full text]  doi: 10.4103/0019-5049.82673
  45. Ko DD, Kang H, Yang SY, Shin HY, Baek CW, Jung YH et al. A comparison of hemodynamic changes after endotracheal intubation by the Optiscope(TM) and the conventional laryngoscope. Korean J Anesthesiol 2012;63(2):130-135. [PubMed] [Free full text]. doi: 10.4097/kjae.2012.63.2.130.
  46. Schalte G, Scheid U, Rex S, Coburn M, Fiedler B, Rossaint R et al. The use of the Airtraq ® optical laryngoscope for routine tracheal intubation in high-risk cardio-surgical patients. BMC Research Notes 2011;4:425. [PubMed] [Free full text] doi: 10.1186/1756-0500-4-425.
  47. Koh JC, Lee JS, Lee YW, Chang CH. Comparison of the laryngeal view during intubation using Airtraq and Macintosh laryngoscopes in patients with cervical spine immobilization and mouth opening limitation. Korean J Anesthesiol 2010;59(5):314-318. [PubMed] [Free full text] doi: 10.4097/kjae.2010.59.5.314.
  48. Kitamura T, Yamada Y, Du HL, Hanaoka K. Efficiency of a new fibreoptic stylet scope in the tracheal intubation. Anesthesiology1999;91:1628-1632. [PubMed] [Free full text]
  49. Kitamura T, Yamada Y, Chinzei M, Du HL, Hanaoka K. Attenuation of haemodynamic responses to tracheal intubation by the StyletScope. Br J Anaesth 2001;86:275-277. [PubMed] [Free full text] doi: 10.1093/bja/86.2.275
  50. Sarkilar G, Sargin M, Saritas TB, Borazan H, Gok F, Kilicaslan A, Otelcioglu S. Hemodynamic responses to endotracheal intubation performed with video and direct laryngoscopy in patients scheduled for major cardiac surgery. Int J Clin Exp Med 2015;8(7):11477-11483. [PubMed] [Free full text]
  51. Bag SK, Hemanth Kumar VR, Krishnaveni N, Ravishankar M, Velraj J, Aruloli M. A comparative study between Truview PCD laryngoscope and Macintosh laryngoscope in viewing glottis opening and ease of intubation: A crossover study. Anesth Essays Res 2014;8:372-6. [PubMed] [Free full text] doi: 10.4103/0259-1162.143152.
  52. Arici S, Karaman S, Dogru S, Karaman T, Tapar H, Ozsoy AZ et al. The McGrath series 5 video laryngoscope versus the Macintosh laryngoscope: a randomized trial in obstetric patients. Turk J Med Sci 2014;44:387-392. [PubMed]
  53. Gupta A, Swetha HR. Comparison between Macintosh laryngoscope and McGrath videolaryngoscope for endotracheal intubation in neurosurgical patients. J Evid Based Med Healthc 2016;3(24):1046-1055. [Free full text]
  54. Tempe DK, Chaudhary K, Diwakar A, Datt V, Virmani S, TomarAS, Mohandas A et al. Comparison of hemodynamic responses to laryngoscopy and intubation with Truview PCD TM, McGrath® and Macintosh laryngoscope in patients undergoing coronary artery bypass grafting: A randomized prospective study. Ann Card Anaesth 2016;19:68-75. [PubMed] [Free full text] doi: 10.4103/0971-9784.173023.
  55. Jakusenko N, Kopeika U, Mihelsons M, Nagobade D, Putnina AV, Pavars A. Comparison of stress response performing endotracheal intubation by direct laryngoscopy, fibreoptic intubation and intubation by the GlideScope laryngoscope. Proceedings of the Latvian Acad Sci. Section B. Natural, Exact and applied. Vol 62(2008) No 4-5(2008) pp 176-181.
  56. Tsai P, Chen B. Hemodynamic responses to endotracheal intubation comparing the Airway Scope®, Glidescope®, and Macintosh laryngoscopes. The internet Journal of Anesthesiology 2009;volume 24 Number 2. [Free full text]
  57. Maassen RLPieters BM , Maathuis B , Serroyen J , Marcus MA. Endotracheal intubation using videolaryngoscopy causes less cardiovascular response compared to classic direct laryngoscopy, in cardiac patients according a standard hospital protocol. Acta Anaesthsiol Belg 2012;63(4):181-186. [PubMed]
  58. Gupta K, Girdhar KK, Anand R, Majgi MS, Gupta PS, Gupta PB. Comparison of haemodynamic response to intubation: Flexible fibreoptic bronchoscope versus bonfils rigid intubation endoscope. Indian J Anaesth 2012 ;56:353-358. [PubMed] [Free full text] doi: 10.4103/0019-5049.100816.
  59. Tomasz Gaszynski, Monika Pietrzyk, Tomasz Szewczyk, and Ewelina Gaszynska, “A Comparison of Performance of Endotracheal Intubation Using the Levitan FPS Optical Stylet or Lary-Flex Videolaryngoscope in Morbidly Obese Patients,”The Scientific World Journal, vol. 2014, Article ID 207591, 9 pages, 2014.